Klin Farmakol Farm. 2026;40(3):164-169 | DOI: 10.36290/far.2026.020

Current options in insulin pump therapy for diabetes

Lucie Radovnická
Interní oddělení, Diabetologické centrum, Masarykova nemocnice v Ústí nad Labem, o. z., Krajská zdravotní, a. s.

Insulin therapy has undergone substantial development over recent decades, evol­ving from insulin pen therapy and conventional insulin pump therapy to automated insulin delivery (AID) systems. A key milestone was the integration of insulin pumps with continuous glucose monitoring, which enabled the transition from sensor-augmented pump therapy to hybrid closed-loop systems. Based on CGM data, current AID systems continuously adjust insulin delivery, thereby improving glycaemic control, increasing time in range, and reducing the risk of hypoglycaemia. In type 1 and 2 diabetes, their benefit has been demonstrated in randomized trials as well as in real-world clinical practice, and these systems have become part of current recommendations. In addition to commercial solutions, open-source systems are also used in practice; they offer greater flexibility but are not officially approved. This article summarizes the development of insulin delivery technologies, the basic principles and types of AID systems, as well as their main limitations and future perspectives. It also briefly addresses the use of these technologies in type 2 diabetes.

Keywords: type 1 diabetes mellitus, type 2 diabetes mellitus, insulin pump, continuous glucose monitoring, automated insulin delivery, hybrid closed-loop system.

Received: March 18, 2026; Revised: May 25, 2026; Accepted: June 1, 2026; Published: October 14, 2026  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
Radovnická L. Current options in insulin pump therapy for diabetes. Klin Farmakol Farm. 2026;40(3):164-169. doi: 10.36290/far.2026.020.
Download citation

References

  1. Pickup J, Keen H, Parsons J, et al. Continuous subcutaneous insulin infusion: an approach to achieving normoglycaemia. Br Med J. 1978;1(6107):204-207. Go to original source...
  2. Pickup J, Keen H, Viberti GC, et al. Continuous subcutaneous insulin infusion in the treatment of diabetes mellitus. Diabetes Care. 1980;3(2):290-300. Go to original source...
  3. Pickup J, Sutton A. Severe hypoglycaemia and glycaemic control in Type 1 diabetes: meta-analysis of multiple daily insulin injections compared with continuous subcutaneous insulin infusion. Diabet Med. 2008;25(7):765-774. Go to original source...
  4. Bergenstal R, Tanborlane WV, Ahmann A, et al. Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes. N Engl J Med. 2010;363(4):311-320. Go to original source...
  5. Bergenstal R, Klonoff DC, Garg SK, et al. Threshold-based insulin-pump interruption for reduction of hypoglycemia. N Engl J Med. 2013;369(3):224-232. Go to original source...
  6. Forlenza G, Li Z, Buckingham B, et al. Predictive Low-Glucose Suspend Reduces Hypoglycemia in Adults, Adolescents, and Children With Type 1 Diabetes in an At-Home Randomized Crossover Study: Results of the PROLOG Trial. Diabetes Care. 2018;41(10):2155-2161. Go to original source...
  7. Brown S, Kovatchev B, Raghinaru D, et al. Six-Month Randomized, Multicenter Trial of Closed-Loop Control in Type 1 Diabetes. N Engl J Med. 2019;381(18):1707-1717. Go to original source...
  8. Breton M, Kanapka L, Beck R, et al. A Randomized Trial of Closed-Loop Control in Children with Type 1 Diabetes. N Engl J Med. 2020;383(9):836-845. Go to original source...
  9. Battelino T, Danne T, Bergenstal R, et al. Clinical Targets for Continuous Glucose Monitoring Data Interpretation: Recommendations From the International Consensus on Time in Range. Diabetes Care. 2019;42(8):1593-1603. Go to original source...
  10. American Diabetes Association Professional Practice Committee for Diabetes. Diabetes Technology: Standards of Care in Diabetes - 2026. Diabetes Care. 2026;49(Suppl 1): S150-S165. Go to original source...
  11. Bergenstal R, Garg S, Weinzimer S, et al. Safety of a Hybrid Closed-Loop Insulin Delivery System in Patients With Type 1 Diabetes. JAMA. 2016;316(13):1407-1408. Go to original source...
  12. van Bon A, Blauw H, Jansen T, et al. Bihormonal fully closed-loop system for the treatment of type 1 diabetes: a real-world multicentre, prospective, single-arm trial in the Netherlands. Lancet Digit Health. 2024;6(4):e272-e280. Go to original source...
  13. Do Q, Haskova A, Radovnicka L, et al. Comparison of Control-IQ and open-source AndroidAPS automated insulin delivery systems in adults with type 1 diabetes: The CODIAC study. Diabetes Obes Metab. 2024;26(1):78-84. Go to original source...
  14. Do Q, Radovnicka L, Haskova A, et al. Switch of Open-Source Automated Insulin Delivery (AID) System-AndroidAPS to Commercially Available AID Systems in Type 1 Diabetes: The Extension of the CODIAC Study. Diabetes Technol Ther. 2025;27(11):928-936. Go to original source...
  15. Levy C, Bailey R, Laffel L, et al. Multicenter Evaluation of Ultra-Rapid Lispro Insulin with Control-IQ Technology in Adults, Adolescents, and Children with Type 1 Diabetes. Diabetes Technol Ther. 2024;26(9):652-660. Go to original source...
  16. Edd S, Castaneda J, Choudhary P, et al. Twelve-month results of the ADAPT randomized controlled trial: Reproducibility and sustainability of advanced hybrid closed-loop therapy outcomes versus conventional therapy in adults with type 1 diabetes. Diabetes Obes Metab. 2023;25(11): 3212-3222. Go to original source...
  17. Shalit R, Minsky N, Laron-Hirsh M, et al. Unannounced Meal Challenges Using an Advanced Hybrid Closed-Loop System. Diabetes Technol Ther. 2023;25(9):579-588. Go to original source...
  18. Scidà G, Corrado A, Abuqwider J, et al. Postprandial Glucose Control With Different Hybrid Closed-Loop Systems According to Type of Meal in Adults With Type 1 Diabetes. J Diabetes Sci Technol. 2025;19(5):1331-1340. Go to original source...
  19. Petruželková L, Neuman V, Plachy L, et al. First Use of Open-Source Automated Insulin Delivery AndroidAPS in Full Closed-Loop Scenario: Pancreas4ALL Randomized Pilot Study. Diabetes Technol Ther. 2023;25(5):315-323. Go to original source...
  20. Haidar A, Legault L, Matteau-Pelletier L, et al. Outpatient overnight glucose control with dual-hormone artificial pancreas, single-hormone artificial pancreas, or conventional insulin pump therapy in children and adolescents with type 1 diabetes: an open-label, randomised controlled trial. Lancet Diabetes Endocrinol. 2015;3(8):595-604. Go to original source...
  21. Do Q, Pehr M, Doležalová B, et al. Účinnost a bezpečnost algoritmu APGO pro automatické dávkování inzulinu: studie RENEW. DMEV. 2025;28(Suppl 1).
  22. Stahl-Pehe A, Shokri-Mashhadi N, Wirth M, et al. Efficacy of automated insulin delivery systems in people with type 1 diabetes: systematic review and network meta-analysis. eCli­ni­calMedicine. 2025;82:103190. Go to original source...
  23. Jendle J, Garg S, Thivolet C, et al. Automated basal insulin delivery versus multiple daily injections in type 1 diabetes: randomized parallel controlled trial. Front Endocrinol. 2025;16:1716587. Go to original source...
  24. de Visser H, Waraich S, Chhabra M, et al. Automated insulin delivery systems and glucose management in children and adolescents with type 1 diabetes: systematic review and meta-analysis. JAMA Pediatr. 2025;179(11):1162-1171. Go to original source...
  25. Pöhlmann J, Smith-Palmer J, Serné E, et al. Systematic review and meta-analysis of real-world evidence on commercially available AID systems in type 1 diabetes. Diabetes Obes Metab. 2026;28(Suppl 1):13-34. Go to original source...
  26. Gargouri I, Charfi H, Belabed W, et al. Precision medicine in type 1 diabetes: Control-IQ vs MiniMed 780G according to characteristics. Diabetes Obes Metab. 2025;27(3):1233-1241. Go to original source...
  27. Santová A, Plachy L, Neuman V, et al. Are all HCL systems the same? Real-life registry study. Front Endocrinol. 2023;14:1283181. Go to original source...
  28. Lee T, Collet C, Bergford S, et al. Automated Insulin Delivery in Women with Pregnancy Complicated by Type 1 Diabetes. N Engl J Med. 2023;389(17):1566-1578. Go to original source...
  29. Rankin D, Hart R, Kimbell B, et al. Rollout of Closed-Loop Technology to Pregnant Women with Type 1 Diabetes: Healthcare Professionals' Views About Potential Challenges and Solutions. Diabetes Technol Ther. 2023;25(4):260-269. Go to original source...
  30. Kudva Y, Raghinaru D, Lum J, et al. A randomized trial of automated insulin delivery in type 2 diabetes. N Engl J Med. 2025;392:1801-1812. Go to original source...
  31. Bhargava A, Bergenstal R, Warren M, et al. Safety and effectiveness of MiniMed 780G AHCL insulin intensification in adults with insulin-requiring type 2 diabetes. Diabetes Technol Ther. 2025;27(5):366-375. Go to original source...
  32. Pasquel F, Davis G, Huffman D, et al. Automated insulin delivery in adults with type 2 diabetes: a nonrandomized clinical trial. JAMA Netw Open. 2025;8(2):e2459348. Go to original source...
  33. Zeng B, Jia H, Yang S, et al. Effectiveness and safety of automated insulin delivery in type 2 diabetes: a meta-analysis. Diabetol Metab Syndr. 2025;17:433. Go to original source...




Clinical Pharmacology and Pharmacy

Madam, Sir,
please be aware that the website on which you intend to enter, not the general public because it contains technical information about medicines, including advertisements relating to medicinal products. This information and communication professionals are solely under §2 of the Act n.40/1995 Coll. Is active persons authorized to prescribe or supply (hereinafter expert).
Take note that if you are not an expert, you run the risk of danger to their health or the health of other persons, if you the obtained information improperly understood or interpreted, and especially advertising which may be part of this site, or whether you used it for self-diagnosis or medical treatment, whether in relation to each other in person or in relation to others.

I declare:

  1. that I have met the above instruction
  2. I'm an expert within the meaning of the Act n.40/1995 Coll. the regulation of advertising, as amended, and I am aware of the risks that would be a person other than the expert input to these sites exhibited


No

Yes

If your statement is not true, please be aware
that brings the risk of danger to their health or the health of others.